Part Number Hot Search : 
ASI10684 8002BT 60150 809LT BGA2717 CA33K032 GP2L26 ASM3P2
Product Description
Full Text Search
 

To Download HM1520 Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 TECHNICAL DATA
PRELIMINARY DATA SHEET
RELATIVE HUMIDITY MODULE
HM1520
Based on the rugged HS1101 humidity sensor, HM1520 is a dedicated humidity transducer designed for measurements at low humidity. Direct measurement of dew point or water concentration can be easily obtained in really cost effective conditions. Direct interface with a micro-controller is made possible with the module's linear voltage output.
Main features
* * * * * * * Tubular form for through wall mounting. Not affected by water condensation Full interchangeability. Typical 1 to 1.6 Volt DC output for 0 to 20% RH at 5 V DC supply. Calibrated, linear voltage for easy electronic interface. Controlled temperature dependency Ratiometric to voltage supply
Humidity sensor specific features
* * * No affected by long period at low humidity values. Patented solid polymer structure. High resistance to chemicals. Fast response time.
*
Maximum ratings
Ratings Supply Voltage (Peak) Operating Range (see graph below) Operating Range (Dew Point) Operating Range in Temperature Symbol Vs RH Td Ta Value 12 0 to 100 -50 to 30 -30 to 60 Unit Vdc % RH C C
Accuracy of HM1520 when used from 1 to 95% RH
10,0 8,0 6,0 4,0 2,0 0,0 -2,0 -4,0 -6,0 -8,0 -10,0
Relative Humidity in % Rh Optimized accuracy range
+/- 3 to 5% RH accuracy range 20 30 40 50 60
+/- 5 to 10% RH accuracy range
0
10
70
80
90
100
Humirel Sensor Device Data / HPC009-20 RevA December 2000
1
TECHNICAL DATA
Characteristics
Characteristics Humidity measuring range / Ta -30 to 60C
(Vs = 5.0Vdc , RL > 1 M otherwise stated)
Symbol
Min
Typ
Max
Unit
Relative Humidity accuracy(1-20%RH) at 23C Relative Humidity accuracy at 55% RH at 23C Voltage supply Nominal output @ RH = 10 % / Ta 25C Current consumption Temperature coefficient(10 to 50 C and 1 to 20% RH) Averaged Sensitivity from 5% to 10 % RH Output Impedance Sink current capability (R L = 15 kohm) Humidity Hysteresis Long term stability Response time (5 to 10% RH, static, @63%)
RH RH RH Vs V out Ic Tcc
mV/%RH
1 +/-2 +/-5 4.75 1.17 5.00 1.24 0.4 < 0.1 +26 70 +/-1.5 0.5 5
99 +/- 3 5.25 1.31 0.8
%RH %RH %RH V V mA %RH / C mV/%RH Ohms A % RH %RH/yr s
Z Is
300
HM1520 Preliminary Specification when used from 1 to 30% RH
2000
Vout= VS ( 5.12 RH + 197) or 1800 * RH = 0.195 Vout / VS - 38.5 1700 with V out in mV, V S in V and RH in %
1900 1600 1500 1400 1300 1200 1100 1000 0 5 10 15 20 25 30 35
Relative Humidity in % RH
*
* * *
Those equations can be used above 30% RH and allow to obtain an over all accuracy as described in page one of this document in all the humidity measuring range Temperature (in the range 0 to 50C) does not affect HM1520 measurement when used from 1 to 30% RH. No temperature compensation is required.
Signal output from 1 to 20% RH at 23C :
RH (%) Vout (mV) RH (%) Vout (mV) 11 1269 0 1 1013 12 1294 2 1038 13 1320 3 1064 14 1346 4 1089 15 1371 5 1115 16 1397 6 1141 17 1422 7 1166 18 1448 8 1192 19 1474 9 1217 20 1499 10 1243
With Vs = 5.0 Volts DC Calibration data are traceable to NIST standards through CETIAT laboratory.
Humirel Sensor Device Data / HPC009-20 RevA December 2000
2
TECHNICAL DATA
IMPORTANT NOTICE :
HM1520 is based on HS1101 capacitive sensor and thus is fully useable on a large range of relative humidity ( 1 to 99% RH). In that range, HM1520 presents a typical accuracy of +/-5% RH at 55% RH.) However HM1520 is a dedicated module for humidity measurements at low Relative Humidity levels. Thus, HM1520 is also well adapted to measure water concentrations (ppm) or low dew points when associated with an ambient temperature probe. HM1520 features an optimized accuracy for water concentration below 6000ppm water or 0C dew point at 23C (equivalent to 20% RH) .
TYPICAL OUTPUT of HM1520 when measuring water concentration at ambient temperature of 23C :
ppm Vout (mV) ppm Vout (mV) 3025 1269 0 275 1013 3300 1294 550 1038 3570 1320 825 1064 3850 1346 1100 1089 4120 1371 1375 1115 4395 1397 1650 1141 4670 1422 1925 1166 4945 1448 2200 1192 5220 1474 2470 1217 5495 1499 2750 1243
ppm water = 10.75 Vout - -10615
with Vout in mVolts
TYPICAL OUTPUT of HM1520 when measuring dew point at ambient temperature of 23C :
C Vout (mV) C Vout (mV) -9 1269 -36 1013 -7.8 1294 -29 1038 -6.8 1320 -24 1064 -5.8 1346 -21 1089 -4.9 1371 -18.5 1115 -4.1 1397 -16 1141 -3.2 1422 -14.5 1166 -2.5 1448 -13 1192 -1.6 1474 -11.5 1217 -1.1 1499 -10 1243
Humirel Sensor Device Data / HPC009-20 RevA December 2000
3
TECHNICAL DATA
Internal Block Diagram
Reference Oscillator L.P. Filter Sensor Oscillator Gain
Vout Vout-->
Package outline HM1520
HM1500
A
DIM A B C D E
MIN 9.70 5.00 52 10.7 100
MAX 10.20 5.50 54 11.1 250
B
DIMENSIONS IN MILLIMETERS D C WIRING WIRE W1 W2 W3 COLOR WHITE BLUE YELLOW FUNCTION GND VOLTAGE OUTPUT
E
W1
W2
W3
The information in this sheet has been carefully reviewed and is believed to be accurate; however, no responsability is assumed for inaccuracies. Furthermore, this information does not convey to the purchaser of such devices any license under the patent rights to the manufacturer. HUMIREL reserves the right to make changes without further notice to any product herein. HUMIREL makes no warranty, representation or guarantee regarding the suitability of its product for any particular purpose, nor does HUMIREL assume any liability arising out of the application or use of any product or circuit and specifically disclaims any and all liability, including without limitation consequential or incidental damages. Typical parameters can and do vary in different applications. All operating parameters, including Typical must be validated for each customer applications by customer's technical experts. HUMIREL does not convey any license under its patent rights nor the rights of others. HUMIREL products are not designed, intended, or authorized for use as components in systems intended for surgical implant into the body, or other application intended to support or sustain life, or for any application in which the failure of the HUMIREL product could create a situation where personal injury or death may occur. Should buyer purchase or use HUMIREL products for any such unintended or unauthorized application, Buyer shall indemnify and hold HUMIREL and its officers, employees, subsidiaries, affiliates and distributors harmless against all claims, costs, damages and expenses, and reasonable attorney fees arising out of, directly or indirectly, any claim of personal injury or death associated with such unintended or unauthorized use, even if such claim alleges that HUMIREL was negligent regarding the design or manufacture of the part .HUMIREL is a registred trade mark of HUMIREL.
Humirel Sensor Device Data / HPC009-20 RevA December 2000
4